HP Indigo 7500 Electrometer Board Test - Page 19

From press UI: Element Activation > Liquids tab, touch the CS, Pump., Cleaning station, Disengage

Page 19 highlights

f. Deactivate the cleaning station oil pump. i. From press UI: Element Activation > Liquids tab, touch the CS Pump. ii. In the Motors Tab, touch the Cleaning station button to deactivate the cleaning station. g. Disengage the PIP drum. From the press UI: (Element Activation > Drums Tab > PIP), press Disengage. h. Stop the motor by clicking the motor button. See step d above for reference. 2. Charged PIP Test Procedure: Turn the electrometer board ON (Via Systab I/O) as follows: a. Get the press to 'Stand-By' mode. Probe Test Procedure 17