TROUBLESHOOTING If any abnormality arises during use, please check the following points SYMPTOMS POSSIBLE CAUSES CORRECTION Unit does not turn on when the START/STOP button is pushed. Batteries have run down. Replace them with two new AA alkaline batteries. Battery polarities Re-insert the batteries have been positioned in the correct incorrectly. positions. EE measurement error Cuff has been placed Wrap the cuff symbol shown on incorrectly. properly so that it is display or the blood positioned correctly. pressure value is Did you talk or move Measure again. displayed excessively during measurement? Keep arm steady low (or high). Shaking of the arm during measurement. with the cuff on. Er error symbol shown Did you move during on display - measure- measurement? ment result exceeds Shaking of the arm measurement range. with the cuff on. Measure again. Do not move during measurement. E1 error symbol shown Air circuit abnormality. Check cuff connection. on display Cuff tube may not be Measure again. plugged into monitor correctly. E2 error symbol shown Inflation pressure Turn the unit off, then on display exceeding 300 mmHg. measure again. E3 error symbol shown Data error. on display Remove and reload the batteries. CA appears on display Recalibration alert. Refer to page 25 for more information. 29 SPECIFICATIONS Power Source: Two 1.5V DC (AA) batteries Measurement Method: Oscillometric Measurement Range: Pressure: 40~250 millimeters Pulse: 40~199 beats/minute mercury (mmHg) Accuracy: Pressure: ±3 mmHg Pulse: ±5% of reading Pressure Sensor: Semi-conductor Inflation: Manual Inflation Deflation: Memory Capacity: Automatic Pressure release valve 60 memories Auto-shut-off: 1 minute after last button operation Operation Environment: Storage Environment: Weight: Temperature: 50°F~104°F (10°C ~40°C) Humidity: 40~85% RH max. Temperature: 14°F~140°F (-10°C~60°C) Humidity: 10~90% RH max. .30 lbs/338 grams (without batteries) Arm Circumference: Standard size adult arm cuff: 9.5"-14.2" (24 - 36 cm) Dimensions: 4.50"(L) x 4.17"(W) x 2.25"(H) 112mm(L) x 106mm(W) x 58mm(H) Type BF Device and cuff are designed to provide special protection against electrical shocks. Note: These specifications are subject to change without notice. 30

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

30
29
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button
is pushed.
Batteries have
run down.
Replace them with
two new AA alkaline
batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries
in the correct
positions.
EE measurement error
symbol shown on
display or the blood
pressure value is
displayed excessively
low (or high).
Cuff has been placed
incorrectly.
Wrap the cuff
properly so that it is
positioned correctly.
Did you talk or move
during measurement?
Measure again.
Keep arm steady
during measurement.
Shaking of the arm
with the cuff on.
Er error symbol shown
on display - measure-
ment result exceeds
measurement range.
Did you move during
measurement?
Shaking of the arm
with the cuff on.
Measure again. Do not
move during
measurement.
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Turn the unit off, then
measure again.
E3 error symbol shown
on display
Data error.
Remove and reload
the batteries.
CA appears on display
Recalibration alert.
Refer to page 25 for
more information.
Power Source:
Two 1.5V DC (AA) batteries
Measurement
Method:
Oscillometric
Measurement Range:
Pressure: 40~250 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Manual Inflation
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C ~40°C)
Humidity: 40~85% RH max.
Storage
Environment:
Temperature: 14°F~140°F (-10°C~60°C)
Humidity: 10~90% RH max.
Weight:
.30 lbs/338 grams (without batteries)
Arm Circumference:
Standard size adult arm cuff:
9.5"–14.2" (24 – 36 cm)
Dimensions:
4.50"(L) x 4.17"(W) x 2.25"(H)
112mm(L) x 106mm(W) x 58mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change without notice.