Invacare 3GRX-CG Owners Manual - Page 58
Caution
View all Invacare 3GRX-CG manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 58 highlights
5 WHEELCHAIR OPERATION 5.5 Charging Batteries ƽ WARNING NEVER attempt to recharge the batteries by attaching cables directly to the battery terminals or clamps. ALWAYS use the recharging plug located on the front of the joystick. DO NOT sit in the wheelchair while charging the batteries. DO NOT attempt to recharge the batteries and operate the power wheelchair at the same time. During use and charging, unsealed batteries will vent hydrogen gas which is explosive in the right concentration with air. CAUTION Always charge new batteries before initial use or battery life will be reduced. For this procedure, refer to FIGURE 11 on page 60. New batteries MUST be fully charged prior to initial use of the wheelchair. As a general rule, batteries should be recharged daily to assure the longest possible life and minimize the required charging time. Plan to recharge the batteries when it is anticipated the wheelchair will not be used for a long period of time. The range per battery charge using recommended batteries should be approximately 5 to 9 hours of typical operation. Extensive use on inclines may substantially reduce per charge mileage. Invacare 3G Storm Series® 58 Part No 1143151