Page 51 highlights
6 Adjustment 6 ADJUSTMENT ƽ WARNING After ANY adjustments, repair or service and before use, make sure that all attaching hardware is tightened securely - otherwise injury or damage may result. While the wheelchair is moving, minimum ground clearance for the front rigging is three inches. If the wheelchair is not moving, the front rigging MUST maintain a minimum of one inch ground clearance - otherwise personal injury and damage may result. CAUTION As with any vehicle, the wheels and tires should be checked periodically for cracks and wear, and should be replaced. 6.1 Footrest Height Adjustment ƽ WARNING While the wheelchair is moving, minimum ground clearance for the front rigging is three inches. If the wheelchair is not moving, the front rigging MUST maintain a minimum of one inch ground clearance otherwise personal injury and damage may result. PH904A and PHAL4A Front Riggings 1. Loosen, but DO NOT remove, the lug bolt and locknut that secure the lower footrest to the footrest support. 2. Reposition the lower footrest to the desired height. 3. Securely tighten the lug bolt and locknut that secure the lower footrest to the footrest support. 4. Repeat STEPS 1-3 for the opposite side of the wheelchair footrest, if necessary. PH904A style front rigging shown. PHAL4A front rigging adjusts the same way. Footrest Support Locknut Lower Footrest Lug Bolt FIGURE 1 Footrest Height Adjustment Part No 1143192 51 Invacare® Adjustable ASBA Seats