Invacare IFX-20MP Owners Manual 2 - Page 48
Step Up/Down
View all Invacare IFX-20MP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 48 highlights
LiNX® Control System 5 Step Up/Down Cruise Control The Latch Drive Timeout period is restarted whenever a subsequent drive command is given. The Latch Drive Timeout is set by the provider. To change the parameter, contact your provider. Turn commands The wheelchair can be steered while in latched driving mode. If a turn command is given, the wheelchair remains in latched driving mode and also responses to the turn command for the duration that the turn command is given. The Latch Drive Timeout period is restarted whenever a turn command is given. When the Latch Drive Timeout expires, the wheelchairs stops. 6.6.1 External stop switch To set up a wheelchair for latched driving, an external stop switch must be fitted to the wheelchair. Ideally, the external stop switch should be highly visible and easily accessible to provide an extra level of safety and security for the user. External stop switch test The external stop switch test checks that the external stop switch is functioning correctly. The test is conducted once per power cycle when: • the wheelchair is powered up in a latched drive mode function or • a latched drive mode function is selected following a non-latched drive mode function. The external stop switch test is indicated by a screen overlay. 1. Press external stop switch to complete test. The wheelchair does not drive until the external stop switch test is completed successfully. 48 60126081-A