Invacare M51P Owners Manual - Page 57
Engaging/Disengaging Motor Release Lever
View all Invacare M51P manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 57 highlights
SECTION 10-TOP SHROUD AND WHEELS Center Hole in Top Shroud Joystick Cable and Connector Base Frame Hook and Loop Strap Top Shroud DETAIL "A" - M61 TOP SHROUD Clip Hook and Loop Strap Front Shroud NOTE: Batteries not shown for clarity. FIGURE 10.1 Removing/Installing the Top Shroud Engaging/Disengaging Motor Release Lever ƽ WARNING DO NOT engage or disengage the motor release lever until the On/Off switch on the joystick is in the Off position. CAUTION Ensure both motor release levers are fully engaged before driving the wheelchair NOTE: For this procedure, refer to FIGURE 10.2 on page 58. NOTE: The motor lock disengagement/engagement allows freewheeling or joystick controlled operation. Freewheeling allows an attendant to maneuver the wheelchair without power. 1. Locate the motor release handles on the motors protruding through the top shroud on the rear of the wheelchair. 2. Perform one of the following: • To Disengage the Motor Release Levers ‐ Pull the motor lock lever towards the rear of the wheelchair (freewheel position). NOTE: This allows the chair to freewheel for pushing, if necessary. NOTE: It may be necessary to rock the wheels slightly until the motor release lever disengages. • To Engage the Motor Release Levers ‐ Push the motor lock handles towards the front of the wheelchair (drive position). NOTE: This allows the motors to drive the wheels. NOTE: It may be necessary to rock the wheels slightly until the motor release lever engages. Part No 1125085 57 Pronto® M51™and M61™with SureStep®