Sony MCC-1000MD Operating Instructions - Page 5

Important EMC notices for use in medical environments, Warning

Page 5 highlights

Important EMC notices for use in medical environments • The MCC-1000MD needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the instructions for use. • The MCC-1000MD is intended for use in a professional healthcare facility environment. • Portable and mobile RF communications equipment, such as cellular phones, can affect the MCC-1000MD. Warning • Portable RF communications equipment should be used no closer than 30 cm (12 inches) to any part of the MCC-1000MD. Otherwise, degradation of the performance of this equipment could result. • If the MCC-1000MD will be used adjacent to or stacked with other equipment, normal operation of the MCC-1000MD under such configurations should be verified via observation. • The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony Corporation, may result in increased emissions or decreased immunity of the MCC-1000MD. List of cables used for EMC test Type of cable Specifications CCMC-SA15 cable 15 m, shielded CCMC-EA05 cable 5 m, shielded Guidance and manufacturer's declaration - electromagnetic emissions The MCC-1000MD is intended for use in the electromagnetic environment specified below. The customer or the user of the MCC-1000MD should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group 1 The MCC-1000MD uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emissions CISPR 11 CISPR 32 Harmonic emissions IEC 61000-3-2 Class B Class A The MCC-1000MD is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Voltage fluctuations/ flicker emissions Complies IEC 61000-3-3 5

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61

5
Important EMC notices for use in medical environments
The MCC-1000MD needs special precautions regarding EMC and needs to be installed and
put into service according to the EMC information provided in the instructions for use.
The MCC-1000MD is intended for use in a professional healthcare facility environment.
Portable and mobile RF communications equipment, such as cellular phones, can affect the
MCC-1000MD.
Warning
Portable RF communications equipment should be used no closer than 30 cm (12 inches) to
any part of the MCC-1000MD. Otherwise, degradation of the performance of this
equipment could result.
If the MCC-1000MD will be used adjacent to or stacked with other equipment, normal
operation of the MCC-1000MD under such configurations should be verified via
observation.
The use of accessories and cables other than those specified, with the exception of
replacement parts sold by Sony Corporation, may result in increased emissions or decreased
immunity of the MCC-1000MD.
List of cables used for EMC test
Type of cable
Specifications
CCMC-SA15 cable
15 m, shielded
CCMC-EA05 cable
5 m, shielded
Guidance and manufacturer’s declaration – electromagnetic emissions
The MCC-1000MD is intended for use in the electromagnetic environment specified below. The
customer or the user of the MCC-1000MD should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The MCC-1000MD uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
CISPR 32
Class B
The MCC-1000MD is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies