Sony MCC-1000MD Operating Instructions - Page 55
Appendices, Troubleshooting, Power, Shooting
View all Sony MCC-1000MD manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 55 highlights
Appendices Troubleshooting Before you call for service, please check the problems and solutions described below. If you cannot solve the problem, contact the nearest authorized dealer. Power Symptom Cause The unit does not power on when you press the 1 (on/standby) switch. The unit is not connected to AC power. All of the LED A system error occurred. indicators on the front panel (except the power indicator) are blinking. Solution Connect AC Power. Remove the power cord and check the camera cable and connection with other devices. If the problem persists, contact Sony service representative. Shooting Symptom Cause Solution No picture is output from the camera. The camera head and CCU are not Check the connection with the camera head. securely connected. The [System] menu's [Region] option Correctly set the [System] menu's [Region] setting (see setting differs from your video page 50). monitor settings. You can also change the [Region] setting as follows. To set to [NTSC]: Press the LOCK button while holding down the MENU and V buttons. To set to [PAL]: Press the LOCK button while holding down the MENU and V buttons. The [System] menu's [SDI Video Format] option setting differs from your video monitor settings. Correctly set the [System] menu's [SDI Video Format] setting (see page 50). Camera picture output is distorted (does not display properly). The camera head and CCU are not securely connected. Check the camera cable connection. Insert the camera cable's connector fully, and turn the connector ring to secure it. 55