Resistance Specifications, Temperature Specifications, Function, Range, Resolution, Accuracy
View all Fluke 287/FVF manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 81 highlights
287/289 Users Manual Resistance Specifications Function Resistance 50 Ω [1][3] 500 Ω [1] 5 kΩ [1] 50 kΩ [1] Range 500 kΩ 5 MΩ 30 MΩ 50 MΩ 50 MΩ up to 100 MΩ Conductance 100 MΩ up to 500 MΩ 50 nS [2] [1] When using the relative mode (REL Q) to compensate for offsets. [2] Add 20 counts above 33 nS in 50 nS range. [3] 289 only. Resolution 0.001 Ω 0.01 Ω 0.0001 kΩ 0.001 kΩ 0.01 kΩ 0.0001 MΩ 0.001 MΩ 0.01 MΩ 0.1 MΩ 0.1 MΩ 0.01 nS Accuracy 0.15 % + 20 0.05 % + 10 0.05 % + 2 0.05 % + 2 0.05 % + 2 0.15 % + 4 1.5 % + 4 1.5 % + 4 3.0 % + 2 8 % + 2 1 % + 10 Temperature Specifications Temperature Resolution Accuracy [1,2] -200 °C to +1350 °C 0.1 °C 1 % + 10 -328 °F to +2462 °F 0.1 °F 1 % + 18 [1] Does not include error of the thermocouple probe. [2] Accuracy specification assumes ambient temperature stable to ±1 °C. For ambient temperature changes of ±5 °C, rated accuracy applies after 2 hours. 72 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664