Fluke 287/FVF Product Manual - Page 84
Frequency Counter Sensitivity, Approximate Current Sensitivity
View all Fluke 287/FVF manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 84 highlights
True-rms Digital Multimeters Detailed Specifications Frequency Counter Sensitivity Input Range Approximate Voltage Sensitivity (rms sine wave) [1] 15 Hz to 100 kHz AC Bandwidth [2] Approximate DC Trigger Levels DC Bandwidth [2] 50 mV 5 mV 1 MHz 5 mV & 20 mV 600 kHz 500 mV 25 mV 1 MHz 20 mV & 60 mV 1 MHz 5 V 0.25 V 700 kHz 1.4 V & 2.0 V 80 kHz 50 V 2.5 V 1 MHz 0.5 V & 6.5 V 1 MHz 500 V 25 V 300 kHz 5 V & 40 V 300 kHz 1000 V 50 V 300 kHz 5 V & 100 V 300 kHz Input Range Approximate Current Sensitivity (rms sine wave) 15 Hz to 10 kHz AC Bandwidth Approximate DC Trigger Levels DC Bandwidth 500 μA 25 μA 100 kHz 5000 μA 250 μA 100 kHz 50 mA 400 mA 2.5 mA 100 kHz NA NA 25 mA 100 kHz 5 A 0.25 A 100 kHz 10 A 1.0 A 100 kHz [1] Maximum input = 10 x range (1000 V maximum, 2 x 107 V-Hz product maximum). Noise at low frequencies and amplitudes may affect accuracy. [2] Typical frequency bandwidth with full scale (or maximum 2 x 107 V-Hz product) rms sine wave. 75 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664