Fluke 287/FVF Product Manual - Page 85
MIN MAX, Recording, and Peak Specifications
View all Fluke 287/FVF manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 85 highlights
287/289 Users Manual MIN MAX, Recording, and Peak Specifications Function Nominal Response Accuracy MIN MAX, Recording Peak Crest Factor 200 ms to 80% (dc function) 350 ms to 80 % (ac function) 250 μS (peak) [1] 350 ms to 80 % Specified accuracy ±12 counts for changes >425 ms in duration in manual range. Specified Accuracy ±40 counts for changes >1.5 s in duration in manual range. Specified accuracy ±100 counts [2] up to 5,000 count (full range) reading. For higher peak reading (to 12,000 counts), specified accuracy ±2 % [3] of reading. For periodic waveforms from 50 to 440 Hz ± (4 % + 1 count). [1] For repetitive peaks; 2.5 ms for single events. Peak not specified for 500 µA DC, 50 mA DC, 5 A DC. [2] 200 counts in 500 mV AC, 500 μA AC, 50 mA AC, 5 A AC. [3] 3 % in 500 mV AC, 500 μA AC, 50 mA AC, 5 A AC. 76 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664