Invacare L-4R Owners Manual - Page 54

Service Indicator Error Codes

Page 54 highlights

14 TROUBLESHOOTING 14.2 Service Indicator Error Codes  For the following table, refer to FIGURE 6.1 on page 25. FLASH DESCRIPTION 1 Battery Low 2 Low Battery Fault 3 High Battery Fault 4 Current Limit Time-out 5 Park Brake Fault 6 Throttle OONAPU 7 Speed Pot Fault 8 Motor Voltage Fault 9 Other error 10 Push/Rollaway Fault MEANING The batteries are running low. • Recharge the batteries. The batteries have run out of charge. • Recharge the batteries. Battery voltage is too high. This may occur if overcharged &/or traveling down a long slope. • If traveling down a slope, reduce your speed to minimize the amount of regenerative charging. • Check the battery and associated connections and wiring. The motor has been exceeding its maximum current rating for too long. This may be due to a faulty motor. • Check the motor and associated connections and wiring. • Turn the controller off, leave for a few minutes and turn back on again. Either a park brake release switch is active or the park brake is faulty. • Check the park brake and associated connections and wiring. • Ensure any associated switches are in their correct positions. The Throttle is out of neutral when turning the controller on. • Ensure the throttle is in neutral when turning the controller on. • The Throttle may require re-calibration. The throttle, speed limit pot, SRW or their associated wiring may be faulty. • Check the throttle and speed pot and associated connections and wiring. The motor or its associated wiring is faulty. • Check the motor and associated connections and wiring. The controller may have an internal fault. • Check all connections & wiring. The scooter has exceeded the Push or Rollaway speed limits. • Turn the controller off and then on again. Lynx™L-3 & Lynx L-4 Scooters 54 Part No. 1143205

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56

14
TROUBLESHOOTING
Lynx™L-3 & Lynx L-4 Scooters
54
Part No. 1143205
14.2
Service Indicator Error Codes
For the following table, refer to
FIGURE 6.1 on page 25.
FLASH
DESCRIPTION
MEANING
1
Battery Low
The batteries are running low.
Recharge the batteries.
2
Low Battery Fault
The batteries have run out of charge.
Recharge the batteries.
3
High Battery Fault
Battery voltage is too high. This may occur if overcharged &/or traveling down a long slope.
If traveling down a slope, reduce your speed to minimize the amount of regenerative charging.
Check the battery and associated connections and wiring.
4
Current Limit Time-out
The motor has been exceeding its maximum current rating for too long. This may be due to a faulty motor.
Check the motor and associated connections and wiring.
Turn the controller off, leave for a few minutes and turn back on again.
5
Park Brake Fault
Either a park brake release switch is active or the park brake is faulty.
Check the park brake and associated connections and wiring.
Ensure any associated switches are in their correct positions.
6
Throttle OONAPU
The Throttle is out of neutral when turning the controller on.
Ensure the throttle is in neutral when turning the controller on.
The Throttle may require re-calibration.
7
Speed Pot Fault
The throttle, speed limit pot, SRW or their associated wiring may be faulty.
Check the throttle and speed pot and associated connections and wiring.
8
Motor Voltage Fault
The motor or its associated wiring is faulty.
Check the motor and associated connections and wiring.
9
Other error
The controller may have an internal fault.
Check all connections & wiring.
10
Push/Rollaway Fault
The scooter has exceeded the Push or Rollaway speed limits.
• Turn the controller off and then on again.