Interfacing (MP225 only) Serial Output The MP225 can interface with computers, printers (GA42), Acquire Data Acquisition software and other RS232 compatible devices via the RS232 cable. The maximum source or sink current available is ± 10mA. Pin Signal Pin 1 device enable (link to pin 5) 6 2 RS232 Tx (signal transmit) 7 3 not used 8 4 not used 9 5 protective ground (0V) Signal not used not used not used not used Data Socket protective ground (0V) signal transmit 5 1 9 6 Type of communication - uni directional, baud rate 2400, data format - 7 data bits : 1 stop bit : even parity 10 n Data is sent to the Data socket at measurement endpoint (manual or automatic), calibration recall and memory recall. n To use the output to monitor continually, you need to select continuous data transfer mode. Recorder Output An output is available for both modes. The polarity of the output is the same as the polarity of the input signal. Approximate Recorder Outputs pH - Output follows display and provides approximately 60 mV output/pH unit. This value varies with changes in electrode slope and temperature. mV - Output follows display, i.e. output varies from -1999 to 1999 mV. Recorder output is not available for temperature. Rec Socket Polarity 0V data output

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15

10
Interfacing
(MP225 only)
Serial Output
The MP225 can interface with computers, printers
(GA42), Acquire Data Acquisition software and other
RS232 compatible devices via the RS232 cable. The
maximum source or sink current available is
±
10mA.
Pin
Signal
Pin
Signal
1
device enable (link to pin 5)
6
not used
2
RS232 Tx (signal transmit)
7
not used
3
not used
8
not used
4
not used
9
not used
5
protective ground (0V)
Data Socket
Type of communication - uni directional, baud rate 2400,
data format - 7 data bits : 1 stop bit : even parity
n
Data is sent to the
Data
socket at measurement
endpoint (manual or automatic), calibration recall and
memory recall.
n
To use the output to monitor continually, you need to
select continuous data transfer mode.
Recorder Output
An output is available for both modes. The polarity
of the output is the same as the polarity of the input
signal.
Approximate Recorder Outputs
pH - Output follows display and provides approxi-
mately 60 mV output/pH unit. This value varies with
changes in electrode slope and temperature.
mV - Output follows display, i.e. output varies from
–1999 to 1999 mV.
Recorder output is not available for temperature.
Rec Socket Polarity
protective
ground (0V)
signal transmit
0V
data output
5
1
9
6