3M MP220 Installation Guide - Page 3

Calibrating - ph meter manual

Page 3 highlights

Calibrating 1. 1-point calibration Example buffer ↓ 2 seconds Use auto endpoint or to freeze the reading. 2. Rinse the electrode and blot dry to return to sample measurement, or continue with step 3 for further cal points 3. Repeat steps 1 and 2 using other buffers for a 2 point (or 3 point - MP225 only) calibration. Electrode Condition Indicator 95.0 - 105.0% electrode in good condition 90.0 - 94.9% electrode needs cleaning 85.0 - 89.9% electrode needs conditioning n When you press Cal the pH buffer you selected for cal 1 (Program Menu) is displayed for 2 seconds, and then the meter starts measuring. If you want to use another pH buffer (choice of 3) press Cal again. Press Cal repeatedly to exit the calibration routine. n The decimal point flashes during calibration measurement. When the electrode output has stabilized the stability indicator appears. Manual endpoint - press Read to endpoint. Auto endpoint - the meter automatically endpoints. n When a calibration has endpointed you can press Mode to display the absolute mV value and temperature of the buffer. The display reverts to the buffer and E0 values, or buffer and slope values for a 2 point (or 3 point - MP225 only) calibration after 2 seconds. 3

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15

3
3.
Repeat steps 1 and 2 using other buffers for a
2 point (or 3 point - MP225 only) calibration.
Electrode Condition Indicator
95.0 - 105.0%
90.0 - 94.9%
85.0 - 89.9%
electrode in
electrode
electrode
good
needs
needs
condition
cleaning
conditioning
n
When you press
Cal
the pH buffer you selected for
cal 1 (Program Menu) is displayed for 2 seconds, and
then the meter starts measuring. If you want to use
another pH buffer (choice of 3) press
Cal
again. Press
Cal
repeatedly to exit the calibration routine.
n
The decimal point flashes during calibration
measurement. When the electrode output
has stabilized the stability indicator appears.
Manual endpoint
- press Read to endpoint.
Auto endpoint
- the meter automatically endpoints.
n
When a calibration has endpointed you can press
Mode
to display the absolute mV value and tempera-
ture of the buffer. The display reverts to the buffer and
E
0
values, or buffer and slope values for a 2 point (or 3
point - MP225 only) calibration after 2 seconds.
buffer
Calibrating
1.
1-point calibration
Example
Use auto endpoint
or
to freeze the reading.
2.
Rinse the electrode and blot dry
to return to sample measurement, or
continue with step 3 for further cal points
2 seconds