Problem Solving Err 1 - offset value (E ) out of range 0 Check correct buffer is used. Check mV reading for pH 7 buffer is 0 ± 30 mV. If it is not, clean or replace the electrode. Err 2 - slope out of range Slope less than 85%, or not calculable. Electrode needs cleaning, conditioning or replacing. Check correct buffers are used. Slope more than 105.0%. Check calibration buffers. NOTE: Where 0 mV/pH unit = 0%, and 59.16 mV/pH unit = 100% (at 25°C) Err 3 - pH buffer outside temperature limits pH buffers must be between 5°C and 50°C for accurate calibration. . Displayed (out of range) Check electrode is connected and immersed in sample. Check wetting cap is removed. Data Entry Errors Date flashes and reverts to previous setting - invalid date entered. Entered temperature value changes to -5.0 or 105.0 - the meter will not accept values outside this range. Dashes displayed instead of buffer value during calibration buffer set up - that buffer has already been selected for a cal point. Cal points must have different values. 11

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15

11
Problem Solving
Err 1
- offset value (E
0
) out of range
Check correct buffer is used.
Check mV reading for pH 7 buffer is 0
±
30 mV. If it
is not, clean or replace the electrode.
Err 2
- slope out of range
Slope less than 85%, or not calculable.
Electrode needs cleaning, conditioning or
replacing. Check correct buffers are used.
Slope more than 105.0%.
Check calibration buffers.
NOTE:
Where 0 mV/pH unit = 0%, and
59.16 mV/pH unit = 100% (at 25
°
C)
Err 3
- pH buffer outside temperature limits
pH buffers must be between 5
°
C and 50
°
C for
accurate calibration.
.
Displayed (out of range)
Check electrode is connected and immersed in
sample. Check wetting cap is removed.
Data Entry Errors
Date flashes and reverts to previous setting - invalid
date entered.
Entered temperature value changes to –5.0 or 105.0
- the meter will not accept values outside this
range.
Dashes displayed instead of buffer value during
calibration buffer set up - that buffer has already
been selected for a cal point. Cal points must have
different values.