Campbell Scientific CS616 CS616 and CS625 Water Content Reflectometers - Page 28

Sensor lead, AM16/32, CR10X

Page 28 highlights

CS616 and CS625 Water Content Reflectometers Sensor lead CS616-1 green CS616-2 green CS616-3 green CS616-1,2,3 orange CS616-4 green CS616-5 green CS616-6 green CS616-4,5,6 orange ... CS616-46 green CS616-47 green CS616-48 green CS616-46,47,48 AM16/32 1H 1L 2H 2L 3H 3L 4H 4L ... 31H 31L 32H 32L RES CLK COM ODD H COM ODD L COM EVEN H COM EVEN L CR10X C1 C2 SE1 SE2 SE3 C3 ;{CR10X} ; *Table 1 Program 01: 60 Execution Interval (seconds) 1: Batt Voltage (P10) ;Measure battery voltage 1: 1 Loc [ Bat_Volt ] 2: Internal Temperature (P17) ;Measure datalogger internal temperature 1: 2 Loc [ DL_Temp ] 3: If time is (P92) ;Every 60 Minutes Start Multiplexer Measurement Loop 1: 0 Minutes (Seconds --) into a 2: 60 Interval (same units as above) 3: 11 Set Flag 1 High 4: If Flag/Port (P91) ;User can start Measurement Loop by setting Flag 1 High 1: 11 Do if Flag 1 is High 2: 30 Then Do 5: Do (P86) 1: 41 ;Set control port 1 high (i.e. Reset Multiplexer) Set Port 1 High 6: Beginning of Loop (P87) 1: 0 Delay 2: 16 Loop Count ;16 loops X 3ea CS616 measured/loop 22

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50

CS616 and CS625 Water Content Reflectometers
Sensor lead
AM16/32
CR10X
CS616-1 green
1H
CS616-2 green
1L
CS616-3 green
2H
CS616-1,2,3 orange
2L
CS616-4 green
3H
CS616-5 green
3L
CS616-6 green
4H
CS616-4,5,6 orange
4L
...
CS616-46 green
31H
CS616-47 green
31L
CS616-48 green
32H
CS616-46,47,48
32L
RES
C1
CLK
C2
COM ODD H
SE1
COM ODD L
SE2
COM EVEN H
SE3
COM EVEN L
C3
;{CR10X}
;
*Table 1 Program
01: 60
Execution Interval (seconds)
1:
Batt Voltage (P10)
;Measure battery voltage
1: 1
Loc [ Bat_Volt
]
2:
Internal Temperature (P17) ;Measure datalogger internal temperature
1: 2
Loc [ DL_Temp
]
3:
If time is (P92)
;Every 60 Minutes Start Multiplexer Measurement Loop
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 11
Set Flag 1 High
4:
If Flag/Port (P91)
;User can start Measurement Loop by setting Flag 1 High
1: 11
Do if Flag 1 is High
2: 30
Then Do
5:
Do (P86)
;Set control port 1 high (i.e. Reset Multiplexer)
1: 41
Set Port 1 High
6:
Beginning of Loop (P87)
;16 loops X 3ea CS616 measured/loop
1: 0
Delay
2: 16
Loop Count
22