Campbell Scientific CS616 CS616 and CS625 Water Content Reflectometers - Page 33
Calibration Equations
View all Campbell Scientific CS616 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 33 highlights
CS616 and CS625 Water Content Reflectometers 7.3 Calibration Equations TABLE 7-1 lists the calibration coefficients derived in the Campbell Scientific soils laboratory. Both linear and quadratic forms are presented. The choice of linear or quadratic forms depends on the expected range of water content and accuracy requirements. These coefficients should provide accurate volumetric water content in mineral soils with bulk electrical conductivity less than 0.5 dS m-1, bulk density less than 1.55 g cm-3, and clay content less than 30%. TABLE 7-1. Standard calibration coefficients for linear and quadratic forms. Linear C0 -0.4677 C1 0.0283 C0 -0.0663 Quadratic C1 -0.0063 C2 0.0007 The linear equation is VWC = -0.4677+0.0283*period . The quadratic equation is VWC = -0.0663 - 0.0063*period + 0.0007*period2 . Period is in microseconds. The result of both calibration equations is volumetric water content on a fractional basis. Multiply by 100 to express in percent volumetric water content. FIGURE 7-3 shows the difference between the linear and quadratic calibration forms over the typical range. A CS616/CS625 output period of 16 microseconds is about 1.2% VWC and 32 microseconds is 44.9%. The linear calibration is within ± 2.7% VWC of the quadratic. The linear calibration underestimates water content at the wet and dry ends of the range and overestimates it by up to about 2.6 % VWC at about 20% VWC. 27