6 General Operation E -04 Increase or decrease the exponent using the U and D buttons. Once the exponent is correct, press Enter to accept it. 6.4.3 ITS-90 PRT and Coefficients The ITS-90 option is for PRTs calibrated and characterized using the International Temperature Scale of 1990 equations. The parameters that appear when ITS-90 is selected are "R0.01", "A", "b", "C", "d", "A4", and "b4". These should be set with the corresponding values that appear on the calibration certificate for the PRT. The parameter "R0.01" takes the triple point of water resistance, often labeled "Rtpw" or "R(273.16K)" on the certificate. Parameters "A", "b", "C", and "d" take the an, bn, cn, and d coefficients where n is a number from 5 to 11. Parameters "A4" and "b4" take the a4 and b4 coefficients or the a5 and b5 coefficients on the certificate. Any ITS-90 parameter of the 1502A that does not have a corresponding coefficient on the PRT's certificate must be set to 0. The following table shows which parameter to set for each of the coefficients that may appear on the certificate. The examples that follow demonstrate how to set the ITS-90 parameters for certain cases. (Note: If the certificate has two sets of coefficients, one set for "zero-power" calibration and one set for 1 mA calibration, use the coefficients for the 1 mA calibration.) Table 1 Matching Certificate Values to 1502A ITS-90 Coefficients 1502A ITS-90 Coefficient A b C d A4 b4 Certificate Value a5, a6, a7, a8, a9, a10, or a11 b5, b6, b7, b8, or b9 c6 or c7 d a4 or a5 b4 or b5 18 Manual Rev. 952801 Hart Scientific MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37

E
-04
Increase or decrease the exponent using the
U
and
D
buttons. Once the ex-
ponent is correct, press
Enter
to accept it.
6.4.3
ITS-90 PRT and Coefficients
The ITS-90 option is for PRTs calibrated and characterized using the
Interna-
tional Temperature Scale of 1990
equations. The parameters that appear
when ITS-90 is selected are “R0.01", ”A", “b”, “C”, “d”, “A4", and ”b4". These
should be set with the corresponding values that appear on the calibration cer-
tificate for the PRT. The parameter “R0.01" takes the triple point of water resis-
tance, often labeled ”Rtpw" or “R(273.16K)” on the certificate. Parameters “A”,
“b”, “C”, and “d” take the a
n
, b
n
,c
n
, and d coefficients where
n
is a number
from 5 to 11. Parameters “A4" and ”b4" take the a4 and b4 coefficients or the
a5 and b5 coefficients on the certificate.
Any ITS-90 parameter of the 1502A
that does not have a corresponding coefficient on the PRT’s certificate must
be set to 0.
The following table shows which parameter to set for each of the coefficients
that may appear on the certificate. The examples that follow demonstrate how
to set the ITS-90 parameters for certain cases. (Note: If the certificate has two
sets of coefficients, one set for “zero-power” calibration and one set for 1 mA
calibration, use the coefficients for the 1 mA calibration.)
18
Manual Rev. 952801
Hart Scientific
6
General Operation
1502A ITS-90 Coefficient
Certificate Value
A
a5, a6, a7, a8, a9, a10, or a11
b
b5, b6, b7, b8, or b9
C
c6 or c7
d
d
A4
a4 or a5
b4
b4 or b5
Table 1
Matching Certificate Values to 1502A ITS-90 Coefficients