Fluke 1502A-156 Product Manual - Page 8
Environmental Conditions
View all Fluke 1502A-156 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 8 highlights
2 Specifications and Environmental Conditions 1Accuracy specifications apply within the recommended operating temperature range. Accuracy limits are increased by a factor of the temperature coefficient outside this range. 2Short-term accuracy includes nonlinearity and noise uncertainties. It does not include drift or calibration uncertainties. 3The temperature range may be limited by the sensor. 4Temperature accuracy is for the 1502A only. It does not include probe uncertainty or probe characterization errors. 2.2 Environmental Conditions Although the instrument has been designed for optimum durability and trouble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and cleaning recommendations can be found in the Maintenance Section of this manual. The instrument operates safely under the following conditions: • temperature range: Absolute 0-55°C (32-131°F); Recommended 16-30°C (61-86°F) • ambient relative humidity: 15-50% • pressure: 75kPa-106kPa • mains voltage within ±10% of nominal • vibrations in the calibration environment should be minimized • altitude does not effect the performance or safety of the unit 4 Manual Rev. 952801 Hart Scientific MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664