6 General Operation Example 1: A PRT was calibrated to ITS-90 and its calibration certificate states values for coefficients Rtpw, a4, b4, a8, and b8. Set the 1502A parameters with values from the certificate as follows. Table 3 Setting Coefficients Rtpw, a8, b8, a4, and b4 1502A Coefficient R0.01 A b C d A4 b4 Certificate Value Rtpw a8 b8 0 0 a4 b4 Example 2: A PRT was calibrated to ITS-90 and its calibration certificate states values for coefficients Rtpw, a5 and b5. Set the 1502A parameters with values from the certificate as follows: Table 2 Setting Coefficients Rtpw, a5, and b5 1502A Coefficient R0.01 A b C d A4 b4 Certificate Value Rtpw a5 b5 0 0 a5 b5 1502A Manual Rev. 952801 19 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37

Example 1:
A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients Rtpw, a4, b4, a8, and b8. Set the 1502A parameters with values
from the certificate as follows.
Example 2:
A PRT was calibrated to ITS-90 and its calibration certificate states values for
coefficients Rtpw, a5 and b5. Set the 1502A parameters with values from the
certificate as follows:
1502A
Manual Rev. 952801
19
6
General Operation
1502A Coefficient
Certificate Value
R0.01
Rtpw
A
a5
b
b5
C
0
d
0
A4
a5
b4
b5
Table 2
Setting Coefficients Rtpw, a5, and b5
1502A Coefficient
Certificate Value
R0.01
Rtpw
A
a8
b
b8
C
0
d
0
A4
a4
b4
b4
Table 3
Setting Coefficients Rtpw, a8, b8, a4, and b4