6 General Operation Example 1: A PRT was calibrated to ITS-90 and its calibration certificate states values for coefficients Rtpw, a4, b4, a8, and b8. Set the 1502A parameters with values from the certificate as follows. Table 3 Setting Coefficients Rtpw, a8, b8, a4, and b4 1502A Coefficient R0.01 A b C d A4 b4 Certificate Value Rtpw a8 b8 0 0 a4 b4 Example 2: A PRT was calibrated to ITS-90 and its calibration certificate states values for coefficients Rtpw, a5 and b5. Set the 1502A parameters with values from the certificate as follows: Table 2 Setting Coefficients Rtpw, a5, and b5 1502A Coefficient R0.01 A b C d A4 b4 Certificate Value Rtpw a5 b5 0 0 a5 b5 1502A Manual Rev. 952801 19 MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664