6 General Operation sponse time. You can set it to any value between 0 and 60 seconds. A value of 0 disables the filter. The default time constant is 4 seconds. To change the filter value, enter the Sample menu. This is done by first pressing the Menu button ("SEt?" appears) then pressing the F/Sample button. The display will briefly indicate "SA Par", then "FILtEr", then the current filter value. Use the U and D buttons to increase or decrease the filter value then press Enter. The next parameter in the menu, the current, will then appear. 6.6 Setting the Current You can select one of two options for the probe excitation current: 1.0 mA (default) or 0.5 mA. The current is set within the Sample menu. Press the Menu button ("SEt?" appears) then press the F/Sample button. The filter value will appear. Press Enter to continue. The display will briefly show "CurrEnt" followed by the value for the current. You can use the Uand Dbuttons to change the value for the current. Press Enter to continue. 6.7 Power Saver The power saver feature is useful for conserving power when operating from a battery. It causes the display to blank after a period of no user activity. The power saver feature can reduce operating current by as much as 100 mA. While the display is blanked a small illuminated dot appears on the left side of the display as an indication that the 1502A is still operating. Pressing any button on the front panel restores the display. You can program the power saver to activate after a specified period of time from 5 minutes to 60 minutes in intervals of 5 minutes. You can also disable the power saver feature completely. The power saver is off by default. The power saver is programmed in the Sample menu. Press the Menu button ("SEt?" appears) then press the Sample menu button. Press Exit twice to skip to the power saver parameter. The display will briefly show "PO SA" followed by the power saver setting. You can use the U and D buttons to change the power saver period (in minutes) or set it to OFF. Press Enter to continue. 22 Manual Rev. 952801 Hart Scientific MyFlukeStore Shop for Fluke products online at: www. .com 1.888.610.7664

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37

sponse time. You can set it to any value between 0 and 60 seconds. A value
of 0 disables the filter. The default time constant is 4 seconds.
To change the filter value, enter the
Sample
menu. This is done by first press-
ing the
Menu
button (“
SEt?
” appears) then pressing the
F/Sample
button.
The display will briefly indicate “
SA Par
”, then “
FILtEr
”, then the current filter
value. Use the
U
and
D
buttons to increase or decrease the filter value then
press
Enter
. The next parameter in the menu, the current, will then appear.
6.6
Setting the Current
You can select one of two options for the probe excitation current: 1.0 mA (de-
fault) or 0.5 mA. The current is set within the
Sample
menu. Press the
Menu
button (“
SEt?
” appears) then press the
F/Sample
button. The filter value will
appear. Press
Enter
to continue. The display will briefly show “
CurrEnt
” fol-
lowed by the value for the current. You can use the
U
and
D
buttons to change
the value for the current. Press
Enter
to continue.
6.7
Power Saver
The power saver feature is useful for conserving power when operating from a
battery. It causes the display to blank after a period of no user activity. The
power saver feature can reduce operating current by as much as 100 mA.
While the display is blanked a small illuminated dot appears on the left side of
the display as an indication that the 1502A is still operating. Pressing any but-
ton on the front panel restores the display. You can program the power saver
to activate after a specified period of time from 5 minutes to 60 minutes in in-
tervals of 5 minutes. You can also disable the power saver feature completely.
The power saver is off by default.
The power saver is programmed in the
Sample
menu. Press the
Menu
button
(“
SEt?
” appears) then press the
Sample
menu button. Press
Exit
twice to
skip to the power saver parameter. The display will briefly show “
PO SA
” fol-
lowed by the power saver setting. You can use the
U
and
D
buttons to
change the power saver period (in minutes) or set it to OFF. Press
Enter
to
continue.