Fluke 190-502/AM Manual - Page 126

Meter Measurements for Fluke 190-xx2, Meter Input Banana Jacks, Meter Functions, General

Page 126 highlights

Fluke 190 Series II Users Manual Meter Measurements for Fluke 190-xx2 The accuracy of all measurements is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. Meter Input (Banana Jacks) Input Coupling DC Frequency Response DC to 3 kHz (-3 dB) Input Impedance 1 MΩ (±1 %)//14 pF (±1.5 pF) Max. Input Voltage Fluke 190-202, -102, -062 1000 V CAT III 600 V CAT IV (For detailed specifications, see "Safety") Max. Input Voltage Fluke 190-502 1000 V CAT II 600 V CAT III (For detailed specifications, see "Safety") Meter Functions Ranging Auto, Manual Modes Normal, Relative General DC Common Mode Rejection (CMRR 100 dB AC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB Ohms (Ω) Ranges 500.0 Ω, 5.000 kΩ, 50.00 kΩ, 500.0 kΩ, 5.000 MΩ, 30.00 MΩ Full Scale Reading 500 Ω to 5 M 5000 counts 30 M 3000 counts Accuracy 0.6 % +5 counts) Measurement Current 0.5 mA to 50 nA, ±20 % decreases with increasing ranges Open Circuit Voltage 4 V Continuity (CONT) Beep 50 Ω (±30 Ω) Measurement Current 0.5 mA, ±20 % Detection of shorts of 1 ms Diode Maximum Voltage Reading 2.8 V Open Circuit Voltage 4 V Accuracy 2 % +5 counts) Measurement Current 0.5 mA, ±20 % 118 MyFlukeStore Shop for Fluke products online at: www. .com 1.877.766.5412

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86
  • 87
  • 88
  • 89
  • 90
  • 91
  • 92
  • 93
  • 94
  • 95
  • 96
  • 97
  • 98
  • 99
  • 100
  • 101
  • 102
  • 103
  • 104
  • 105
  • 106
  • 107
  • 108
  • 109
  • 110
  • 111
  • 112
  • 113
  • 114
  • 115
  • 116
  • 117
  • 118
  • 119
  • 120
  • 121
  • 122
  • 123
  • 124
  • 125
  • 126
  • 127
  • 128
  • 129
  • 130
  • 131
  • 132
  • 133
  • 134
  • 135
  • 136
  • 137
  • 138
  • 139
  • 140
  • 141
  • 142
  • 143
  • 144
  • 145
  • 146
  • 147
  • 148
  • 149
  • 150
  • 151
  • 152
  • 153
  • 154
  • 155
  • 156
  • 157
  • 158

Fluke 190 Series II
Users Manual
118
Meter Measurements for Fluke 190-xx2
The accuracy of all measurements is within
±
(% of
reading + number of counts) from 18
°
C to 28
°
C.
Add 0.1x (specific accuracy) for each
°
C below 18
°
C or
above 28
°
C.
Meter Input (Banana Jacks)
Input Coupling
..............................................................
DC
Frequency Response
.........................
DC to 3 kHz (-3 dB)
Input Impedance
................
1 M
Ω
(±1 %)//14 pF (±1.5 pF)
Max. Input Voltage Fluke 190-202, -102, -062:
...................................................................
1000 V CAT III
600 V CAT IV
(For detailed specifications, see “Safety”)
Max. Input Voltage Fluke 190-502:
....................................................................
1000 V CAT II
600 V CAT III
(For detailed specifications, see “Safety”)
Meter Functions
Ranging
.......................................................
Auto, Manual
Modes
.....................................................
Normal, Relative
General
DC Common Mode Rejection (CMRR)
..................
>100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz
......
>60 dB
Ohms (
Ω
)
Ranges
..................................
500.0
Ω
, 5.000 k
Ω
, 50.00 k
Ω
,
500.0 k
Ω
, 5.000 M
Ω
, 30.00 M
Ω
Full Scale Reading
500
Ω
to 5 M
Ω
...............................................
5000 counts
30 M
Ω
...........................................................
3000 counts
Accuracy
...............................................
±
(0.6 % +5 counts)
Measurement Current
...................
0.5 mA to 50 nA,
±
20 %
decreases with increasing ranges
Open Circuit Voltage
....................................................
<4 V
Continuity (CONT)
Beep
...........................................................
<50
Ω
(
±
30
Ω
)
Measurement Current
..................................
0.5 mA,
±
20 %
Detection of shorts of
................................................
1 ms
Diode
Maximum Voltage Reading
......................................
2.8 V
Open Circuit Voltage
.................................................
<4 V
Accuracy
...............................................
±
(2 % +5 counts)
Measurement Current
...............................
0.5 mA,
±
20 %
Shop for Fluke products online at:
1.877.766.5412
www.
MyFlukeStore
.com