Fluke 576 FE 576 Users Manual - Page 29
Symptom, Display Code, Problem, Action, Auto/Allset to All
View all Fluke 576 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 29 highlights
576 Symptom Display Code Problem Action Photo quality Light is insufficient for photo Use Photoflash diminished with without flash always on (Dip switch available light. "Auto/All"set to "All" (Abrupt transitions or provide more between color shades) ambient light. Connection of unit Wrong com port selected Select correct com to PC software or correct com port in use. or disconnect other does not happen device from correct when software started com port No picture in More than one video Right mouse click "Live" mode source is installed on the "Live" mode window window. Choose correct "Video source" in the menu. Unit indicates The light is too bright for Go to next location "Use flash!" in a photo or reduce brightness. bright light Unit leaves LOG The last position has been used If another recording after a photo and to record photo and data. of the data and data recorded. photo is required, then push Enter button and use arrow keys to re-record a position. Photo Mem! All memory for storing photos 1. Download existing has been used. photos and data and clear memory. - or - 2. If you don't mind losing photos in the unit, you can go to Setup in the unit, push the button until "Clear photo" appears in the display and press Enter. 29