Fluke 576 FE 576 Users Manual - Page 32
Specifications
View all Fluke 576 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 32 highlights
576 Specifications Temp. Range Display Resolution Accuracy (Infrared) at 25°C (77°F) ambient temperature Ambient Derating Optical Resolution (Standard Focus) Optical Resolution (Close Focus) Accuracy (Thermocouple K & J) Accuracy (Thermistor) Repeatability (Infrared) Response Time (95%) Hot Spot Detection (30%) Spectral Range Ambient Operating Range Storage Temperature Relative Humidity Analog Output (optional cable needed) Digital Output Power Dimensions Tripod Mount 32 - 30 to 900°C (- 25 to 1600°F) 0.1°C (0.2°F) ± 0.75% of reading or ± 1 K (± 1.5°F), whichever is greater ± 2°C (± 4°F) for targets below -5°C (23°F) < 0.05K/K or < 0.05%/K, whichever is greater at + 25°C (77°F) ± 25° (± 45°F) 60:1 (19mm spot size at 1.15 m) (0.75in. spot size at 3.8 feet) 50:1 ( 6mm spot size at 0.3 m) (0.24in. spot size at 0.98 feet) ± 2°C or ± 0.75%, whichever is greater -30 to 0°C (-22 to 32°F) ± 0.6K 0 to 70°C (32 to 158°F) ± 0.4K 70 to 100°C (158 to 212°F) ±1K 100 to 120°C (212 to 248°F) ±1.5K ± 0.5% of reading or ± 0.5°C (1°F), whichever is greater, ± 1°C (± 2°F) for targets below -5°C (23°F) 250 mSec 85 mSec 8 to 14 µm 0 to 50°C (32 to 122°F) -20 to 50°C (-4 to 122°F) without batteries 10 to 90% at 30°C (86°F), non condensing 1 mV/°C (°F) USB 1.1 2 x 1.5 V Alkaline Type AA 240 x 170 x 50 mm (7.9 x 6.7 x 2 inches) 1/4"-20 UNC