Fluke 725Ex FE 725ex Users Manual - Page 32
Measuring Temperature, Using Thermocouples
View all Fluke 725Ex manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 32 highlights
725Ex Users Manual Measuring Temperature Using Thermocouples The Calibrator supports twelve standard thermocouples, including types E, N, J, K, T, B, R, S, L, XK, BP, and U. Table 5 summarizes the ranges and characteristics of the supported thermocouples. To measure temperature using a thermocouple, proceed as follows: 1. Attach the thermocouple leads to the appropriate TC miniplug, then to the TC input/output as shown in Figure 10. WCaution One thermocouple pin is wider than the other. To avoid possible damage to Calibrator or to equipment under test do not try to force a miniplug in the wrong polarization. Note If the Calibrator and the thermocouple plug are at different temperatures, wait one minute or more for the connector temperature to stabilize after plugging the miniplug into the TC input/output. M 2. If necessary, press for MEASURE mode. 3. Press T for the TC display. Continue pressing this key to select the desired thermocouple type. If necessary, toggle between °C or °F temperature units by pressing D. 22