Fluke 725Ex FE 725ex Users Manual - Page 50
Auto Ramping the Output, Storing and Recalling Setups
View all Fluke 725Ex manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 50 highlights
725Ex Users Manual Auto Ramping the Output Auto ramping gives the ability to continuously apply a varying stimulus from the Calibrator to a transmitter, while your hands remain free to test the response of the transmitter. When L is pressed, the Calibrator produces a continuously repeating 0 % - 100 % - 0 % ramp in a choice of three ramp waveforms: • E 0 % - 100 % - 0 % 40-second smooth ramp • P 0 % - 100 % - 0 % 15-second smooth ramp • N 0 % - 100 % - 0 % Stair-step ramp in 25 % steps, pausing 5 seconds at each step. Steps are listed in Table 7. To exit ramping, press any button. Table 7. mA Step Values Step 4 to 20 mA 0 % 25 % 50 % 75 % 100 % 4.000 8.000 12.000 16.000 20.000 Storing and Recalling Setups Store up to eight settings in a nonvolatile memory and recall the settings for later use. A low battery condition or a battery change does not jeopardize the stored settings. Proceed as follows: 1. After creating a calibrator setup, press S. In the display, the memory locations appear. 2. Press Y or Z to select locations one through eight. An underscore appears below the selected memory location. 3. Press S until the memory number disappears then reappears. The setup is stored. To recall setups, proceed as follows. 1. Press Q. The memory locations appear on the display. 2. Press Y or Z to select the appropriate location and press Q. 40