Fluke 87V FE 83V & 87V Users Manual - Page 36
Fluke 87V Manual
View all Fluke 87V manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 36 highlights
80 Series V Users Manual Table 6. Functions and Trigger Levels for Frequency Measurements Function K K mL L L L L Range 6 V, 60 V, 600 V, 1000 V 600 mV 600 mV 6 V 60 V 600 V 1000 V Approximate Trigger Level ± 5 % of scale ± 30 mV 40 mV 1.7 V 4 V 40 V 100 V Typical Application Most signals. High-frequency 5 V logic signals. (The dc-coupling of the L function can attenuate high-frequency logic signals, reducing their amplitude enough to interfere with triggering.) Refer to the measurement tips given before this table. 5 V logic signals (TTL). Automotive switching signals. Refer to the measurement tips given before this table. R e E Gd Frequency counter characteristics are not available or specified for these functions. \ All ranges ± 5 % of scale AC current signals. µAF 600 µA, 6000 µA 30 µA , 300 µA Refer to the measurement tips given before this table. ^ 60 mA, 400 mA 3.0 mA , 30 mA AF 6 A, 10 A .30 A, 3.0 A 28