Table 19. MIN MAX Recording Specifications, Model, Nominal Response, Accuracy
View all Fluke 87V manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 60 highlights
80 Series V Users Manual Table 19. MIN MAX Recording Specifications Model Nominal Response Accuracy 83 100 ms to 80 % Specified accuracy ± 12 counts for changes > 200 ms in duration (± 40 counts in ac with beeper on) 87 100 ms to 80 % (dc functions) 120 ms to 80 % (ac functions) 250 µs (peak) (Model 87 only)1 Specified accuracy ± 12 counts for changes > 200 ms in duration Specified accuracy ± 40 counts for changes > 350 ms and inputs > 25 % of range Specified accuracy ± 100 counts for changes > 250 µs in duration (add ± 100 counts for readings over 6000 counts) (add ± 100 counts for readings in Low Pass mode) 1. For repetitive peaks: 1 ms for single events. 52