HP Indigo 7800 Leak Valves Flow Calibration -- CA393-08710 Rev 01 - Page 9

HP Indigo 7800 Manual

Page 9 highlights

4. Connect the CE laptop to the feeder. Use a serial cable (with a USB-serial cable adaptor if necessary). NOTE: Alternatively, you can perform the calibration directly through Print Care. 5. Open the folder C:\Program Files\FDT, and double-click feederdiag.exe to start the FDT. Make sure that the reset feeder on program start option is selected in the initial FDT startup screen. NOTE: See Feeder Diagnostic Tool (CA393-02630) for further instructions on operating the FDT. 6. Turn on the calibration mode by clicking Turn calibration module On/Off in the Manual operations screen. To turn off the diagnostic mode, click Toggle. 7. Load 90 gsm gloss substrate into the feeder drawers 2 and 3. 8. In the FDT, go to the Air pressure calibration area to set the parameters and perform the calibration. Figure 2-7 FDT - Air pressure calibration area a. In the GSM drop-down list, select the substrate weight 090. b. In the Drawer field, enter a drawer number (either 2 or 3). c. Click Airflow and view the measurement results on the voltmeter. Calibrating the flow of the air flow leak valve 7

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12

4.
Connect the CE laptop to the feeder. Use a serial cable (with a USB-serial cable adaptor if necessary).
NOTE:
Alternatively, you can perform the calibration directly through Print Care.
5.
Open the folder
C:\Program Files\FDT
, and double-click
feederdiag.exe
to start the FDT.
Make sure that the
reset feeder on program start
option is selected in the initial FDT startup screen.
NOTE:
See
Feeder Diagnostic Tool
(CA393-02630) for further instructions on operating the FDT.
6.
Turn on the calibration mode by clicking
Turn calibration module On/Off
in the
Manual operations
screen.
To turn off the diagnostic mode, click
Toggle
.
7.
Load 90 gsm gloss substrate into the feeder drawers 2 and 3.
8.
In the FDT, go to the
Air pressure calibration area
to set the parameters and perform the calibration.
Figure 2-7
FDT — Air pressure calibration area
a.
In the
GSM
drop-down list, select the substrate weight
090
.
b.
In the
Drawer
field, enter a drawer number (either 2 or 3).
c.
Click
Airflow
and view the measurement results on the voltmeter.
Calibrating the flow of the air flow leak valve
7