HP Latex 3600 Site Preparation Guide - Page 12
Configuration 2: 200–240 V line-to-line three-phase configuration
View all HP Latex 3600 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 12 highlights
Configuration 2: 200-240 V line-to-line three-phase configuration Table 2-5 Three-phase specifications Number of power wires) 4 (L1/L2/L3/PE) Input voltage (line-to-line) 200-240V (±10%) Input frequency 50/60 Hz Power consumption (typical) 9-11 kW Maximum load current (per phase) 56 A Table 2-6 Branch circuit breaker specifications Three-phase 3 poles, 63/70 A Table 2-7 AC three-phase power cable specifications Configuration 4 wires, L1/L2/L3/PE Wire Strained Cu minimum, 10 mm² or 6 AWG Terminals Lines, ferrule terminals, PE, M8 ring terminal External diameter range 22.0-33.0 mm Configuration 3: 380-415 V line-to-line three-phase configuration with single-phase control Table 2-8 Specifications Three-phase line Single-phase control Number of power wires 5 (L1/L2/L3/N/PE) 3 (L/N/PE) Input voltage (line-to-line) 380-415 V~ (±10%) 100-240 V (±10%) Input frequency 50/60 Hz 50/60 Hz Power consumption (typical) 9-11 kW 0.5 kW Maximum load current (per phase) 35 A 10 A Table 2-9 Branch circuit breaker specifications Three-phase 4 poles, 40/50 A Two-phase control 2 poles, 15/16/20 A Table 2-10 AC Power cable specifications Three-phase line Configuration 5 wires, L1/L2/L3/N/PE Wire Strained Cu minimum, 10 mm² or 8 AWG Single-phase line 3 wires, L/N/PE Strained Cu minimum, 1.5 mm² or 16 AWG 8 Chapter 2 Site preparation ENWW