Xerox 790DP Fiery SI Color Server User Guide - Page 85
In the PowerBook Setup Control Panel, select Normal for Modem Controls., MAIN MENU, CALIBRATION
![]() |
UPC - 042215474375
View all Xerox 790DP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 85 highlights
66-7 Understanding calibration 6. Configure the densitometer to factory defaults and save the settings. (See the densitometer manual for details.) Factory default settings for the X-Rite DTP32 are: Page 1 Page 2 Page 3 Page 4 Page 5 LANG (ENG) BAUD (9600) DPT (on) DEF (off ) M/M (max) TONE (loud) HAND (xon) SEP (spc) X10 (off ) LOCK (off ) TYPE (ser) AXMT (on) DLIM (crlf ) DAP (off ) 7. Turn on the PowerBook. 8. In the PowerBook Setup Control Panel, select Normal for Modem Controls. 9. Calibrate the densitometer, as described below. TO CALIBRATE THE X-RITE DTP32: 1. On the X-Rite DTP32, simultaneously press the two buttons marked MENU. The words MAIN MENU appear in the display. 2. Press the p1 key until p2 appears. 3. Press the cal key. CALIBRATION appears in the display. 4. Press the den key. CALIBRATING TRANSMISSION is displayed, followed by the words INSERT CAL STRIP. 5. Position the strip guide at 5. Strip guide 6. Insert the arrow end of the Auto-Cal strip into the entrance of the X-Rite DTP32 until the roller starts pulling the strip.