Xerox 790DP Fiery SI Color Server User Guide - Page 86
Measuring values with a DTP32 densitometer
UPC - 042215474375
View all Xerox 790DP manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 86 highlights
66-8 Color Calibration READING appears momentarily, followed by the density values and CALIBRATION OK. If UNRECOGNIZABLE STRIP appears, try cleaning the strip (see your X-Rite DTP32 Operating Manual). 7. Recalibrate the densitometer at least once month. Measuring values with a DTP32 densitometer You can connect a DTP32 densitometer to the Fiery SI or to your Macintosh or PC if you are using Fiery XJ Print Calibrator. The densitometer scans the density of the first patch, automatically advances to the next patch, and automatically transmits the measured values to the Fiery SI or Fiery XJ Print Calibrator. The DTP densitometer has an adjustable strip guide to the right of the strip entrance. After each column is scanned, the density measurements are used to create a Measurements file. In general, valid measurements for all patches fall within the range of -0.05 to 3.05. TO MEASURE CALIBRATION PATCHES WITH AN X-RITE DTP32: 1. When prompted on the Control Panel or Fiery XJ Print Calibrator, position the pointer on the strip guide to 15 and click OK on the Control Panel or Fiery XJ Print Calibrator to begin measuring. First, you'll measure the cyan column. Slide the paper guide left or right until the arrow points to 15.