Typical Product Parameters - 5 model #
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 14 highlights
SECTION 4-TYPICAL PRODUCT PARAMETERS Flow Range: Low Flow Alarm: Average Power Consumption: Pressure Relief Mechanism Operational at: Change in maximum recommended flow when back pressure of 7kPa is applied: Filters: Safety System: Width: Height: Depth: Weight: Shipping Weight: Platinum 5 and XL - 0.5 to 5 L/min For flowrates less than 1 L/min, we recommend the use of the Invacare Pediatric Flowmeter IRCPF16 Accessory. Platinum 10 - 2 L/min to 10 L/min Platinum 5 - 0 L/min to 0.5 L/min, Platinum 10 - 0 L/min to 1 L/min Rapid Audible Alarm Beeping (No Accessories Connected) Platinum 5 and XL: 400W Platinum 10: 585W 35 psi ± 3.5 psi (241 kPa ± 24.1 kPa) 0.7 L/min Cabinet (2), Outlet HEPA and Compressor Inlet. Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm w/compressor shutdown. Low Pressure Alarm w/compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System (SensO2 Model) Low Flow Alarm. 18 3/8 in ± 3/8 in (46.7 cm ± 1 cm) 26 3/8 in ± 3/8 in (67.0 cm ± 1 cm) 14 3/8 in ± 3/8 in (36.5 cm ± 1 cm) Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± 1 kg) Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg) Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± 1 kg) Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg) Platinum™ Series 14 Part No 1118353