Invacare IRC5LX Owners Manual - Page 26
Troubleshooting, Guide - parts
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 26 highlights
SECTION 7-TROUBLESHOOTING GUIDE SECTION 7-TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: SHORT BEEPS, LONG PAUSE Main Power Loss: 1. Power cord NOT plugged in. 1. Insert plug into outlet. 2. No power at outlet. 2. Inspect house circuit break- Concentrator ers or fuses. If problem recurs, NOT operating, use a different outlet. power switch ON. 3. Tripped circuit breaker. 3. Call Service Provider. Beep......Beep...... Alarm: System Failure: 1a. Remove and clean cabinet CONTINUOUS 1. Unit overheating due to filters. blocked air intake. Concentrator NOT operating, Power Switch ON. 2. Insufficient power at 1b. Move oxygen concentrator at least three inches away from walls, draperies or furniture. 2. DO NOT use extension Beeeeeeeep.... outlet. cords. Move to another electrical outlet or circuit. 3. Internal repairs 3. Call Service Provider. required Alarm: RAPID Beep....Beep... Beep...Beep Low Flow Alarm: 1. Inspect for kinks or 1. Kinked or blocked tub- blockages. Correct, clean or ing, cannula or humidifier. replace item. Once corrected, turn power OFF for 60 seconds and then turn power back ON. 2. Flowmeter set too low: 2. Reset flowmeter to pre- 0.5 L/min (or less) for scribed flowrate. If prescribed Platinum 5; 1.0 L/min (or flowrate is below 0.5 L/min, less) for Platinum 10. use the Pediatric Flowmeter IRCPF16 (Only with Platinum 5 and XL). Platinum™ Series 26 Part No 1118353