Invacare IRC5LX Owners Manual - Page 20
Flowrate
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 20 highlights
SECTION 5-OPERATING INSTRUCTIONS Flowrate NOTE: For this procedure, refer to FIGURE 5.6 on page 20. 1. Turn the flowrate knob to the setting prescribed by your physician or therapist. CAUTION Never set the flow greater than 5 L/min on the IRC5. Never set the flow greater than 10L/min on the IRC10. ƽ WARNING DO NOT change the L/min setting on the flowmeter unless a change has been prescribed by your physician or therapist. NOTE: To properly read the flowmeter, locate the prescribed flowrate line on the flowmeter. Next, turn the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed. 2. If the flowrate on the flowmeter ever falls below 0.5 L/min for Platinum 5 (1 L/min for Platinum 10) for more than about one minute, the LOW FLOW alarm will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective humidifier bottle. After rated flow is restored, more than 0.5 L/min for Platinum 5 (1 L/min for Platinum 10), the LOW FLOW audible alarm will go off. NOTE: The use of some accessories such as the pediatric flowstand and the HomeFill compressor will deactivate the Low Flow Alarm. Flow Knob Platinum™ Series Ball FIGURE 5.6 Flowrate 20 Part No 1118353