Invacare IRC5LX Owners Manual - Page 25
Invacare IRC5LX Manual
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 25 highlights
MODEL NO. I RC Part No 1118353 FIGURE 6.2 Preventive Maintenance Record 25 PREVENTIVE MAINTENANCE RECORD INVACARE PLATINUM OXYGEN CONCENTRATOR ON EACH INSPECTION Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min. Flow Rate EVERY 4,380 HOURS FOR IRC5LX (EVERY 2,160 HOURS FOR IRC10LX) Check Oxygen Concentration (UNITS WITHOUT SensO 2 ) EVERY 26,280 HOURS OR 3 YEARS, WHICHEVER COMES FIRST Check Oxygen Concentration (UNITS WITH SensO 2) DURING PREVENTIVE MAINTENANCE SCHEDULE, OR BETWEEN PATIENTS Clean/Replace Cabinet Filter(s) Check Outlet HEPA Filter* Check Compressor Inlet Filter* Check Power Loss Alarm AS REQUIRED Clean Heat Exchanger Replace Exhaust Muffler Rebuild Top End of Compressor *NOTE: Refer to Preventive Maintenance section of Service Manual. NOTE: 2,160 hours are equivalent to usage 24 hours per day, for 90 days. 4,380 hours are equivalent to usage 24 hours per day, 7 days per week, for 6 months. 26,280 hours are equivalent to usage 24 hours per day, 7 days per week, for 3 years. SECTION 6-MAINTENANCE SERIAL NO. Platinum™ Series