Invacare IRC5LX Owners Manual - Page 24
Cleaning the Cabinet, Cleaning the Humidifier
View all Invacare IRC5LX manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 24 highlights
SECTION 6-MAINTENANCE 2. Clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly. 3. Dry the filters thoroughly before reinstallation. Cabinet Filter (located on both sides of concentrator) FIGURE 6.1 Cleaning the Cabinet Filter Cleaning the Cabinet ƽ WARNING UNPLUG the concentrator when cleaning. To avoid electrical shock, DO NOT remove cabinet. Clean the cabinet with a mild household cleaner and non‐abrasive cloth or sponge. Cleaning the Humidifier NOTE: To clean the oxygen humidifier, follow the instructions provided by the manufacturer. If none are provided, follow these STEPS: 1. Clean the humidifier every day. 2. Wash it in soapy water and rinse it with a solution of ten parts water and one part vinegar. 3. Rinse thoroughly with hot water and refill with distilled water to the level shown on the humidifier. Platinum™ Series 24 Part No 1118353