SECTION 6-MAINTENANCE SECTION 6-MAINTENANCE ƽ WARNING The Invacare concentrators are specifically designed to minimize routine preventive maintenance. Only qualified personnel should perform preventive maintenance on the concentrator. Power should be disconnected before beginning preventive maintenance on the Invacare concentrator. NOTE: At a minimum, preventive maintenance MUST be performed according to the maintenance record guidelines. In places with high dust or soot levels, maintenance may need to be performed more often. Refer to Preventive Maintenance Record on page 25. Routine Maintenance NOTE: The following routine maintenance should be performed more frequently. See recommended intervals in each section and preventive maintenance record. Cleaning the Cabinet Filter CAUTION DO NOT operate the concentrator without the filter installed. NOTE: For this procedure, refer to FIGURE 6.1 on page 24. NOTE: There are two cabinet filters one located on each side of the cabinet. 1. Remove each filter and clean at least once a week depending on environmental conditions. NOTE: Environmental conditions that may require more frequent cleaning of the filters include but are not limited to: high dust, smoking, air pollutants, etc. Part No 1118353 23 Platinum™ Series

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32

SECTION 6—MAINTENANCE
Part No 1118353
23
Platinum
Series
SECTION 6—MAINTENANCE
±
WARNING
The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. Only qualified
personnel should perform preventive maintenance on the
concentrator.
Power should be disconnected before beginning preventive
maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. In places
with high dust or soot levels, maintenance may need to be performed
more often. Refer to Preventive Maintenance Record
on page 25.
Routine Maintenance
NOTE: The following routine maintenance should be performed
more frequently. See recommended intervals in each section and
preventive maintenance record.
Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 24.
NOTE: There are two cabinet filters one located on each side of the
cabinet.
1.
Remove each filter and clean at least once a week
depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust,
smoking, air pollutants, etc.