Invacare MA95Z Owners Manual - Page 13
Performance Specifications, Mattress Weight Capacity, Pressure Zone, Patient Comfort Control
View all Invacare MA95Z manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 13 highlights
Performance Specifications MA90Z MA95Z Mattress Weight Capacity Standard - 36 inch: 350 lbs Bariatric - 42 Inch: 650 lbs Bariatric - 48 Inch: 1000 lbs Pressure Zone Maximum Flow: 50 ± 15 LPM 1275 ± 100 LPM Maximum Pressure: 35 ± 5 mmHg Maximum Flow Timer: 30 minutes 15 minutes Support Surface Inflation Time: 5-10 minutes 25-60 seconds Patient Comfort Control Pressures: Soft Pressure: 8 ± 4 mmHg Firm Pressure: 32 ± 4 mmHg Rotation Time: 10, 20, 30, 60 5 minutes-4 hrs minutes Rotation Angle: Patient Contact: Overview 0-30° ± 5° 0-40° ± 5° Control unit and mattress have Latex-Free components 1148139-H 13
Overview
Performance Specifications
MA90Z
MA95Z
Mattress Weight Capacity
Standard — 36
inch:
350 lbs
Bariatric — 42
Inch:
650 lbs
Bariatric — 48
Inch:
1000 lbs
Pressure Zone
Maximum Flow:
50 ± 15 LPM
1275 ± 100 LPM
Maximum
Pressure:
35 ± 5 mmHg
Maximum Flow
Timer:
30 minutes
15 minutes
Support Surface
Inflation Time:
5–10 minutes
25–60 seconds
Patient Comfort Control Pressures:
Soft Pressure:
8 ± 4 mmHg
Firm Pressure:
32 ± 4 mmHg
Rotation Time:
10, 20, 30, 60
minutes
5 minutes–4 hrs
Rotation Angle:
0–30° ± 5°
0–40° ± 5°
Patient Contact:
Control unit and mattress have
Latex-Free components
1148139-H
13