Invacare MA95Z Owners Manual - Page 37

Problem, Cause, Solution

Page 37 highlights

Problem Fowler Transmitter not working Alarm-Power Failure Alarm-Hose Disconnected Cause Low voltage in battery Malfunctioning sensor Sensor not synchronized to power unit Power cord disconnected No power in the power source Power outage Blown fuse Mattress hose disconnected Troubleshooting Solution Replace battery Replace fowler Refer to Folwer transmitter setup in this manual Connect power cord to the power source Check power source has power and turn it On Wait till the power source has power Replace blown fuse with an equivalent fuse Connect hose connectors and lock them in place 1148139-H 37

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

Troubleshooting
Problem
Cause
Solution
Low voltage in battery
Replace battery
Malfunctioning sensor
Replace fowler
Fowler Transmitter not working
Sensor not synchronized to power unit
Refer to Folwer transmitter setup in this
manual
Power cord disconnected
Connect power cord to the power source
No power in the power source
Check power source has power and turn
it On
Power outage
Wait till the power source has power
Alarm—Power Failure
Blown fuse
Replace blown fuse with an equivalent
fuse
Alarm—Hose Disconnected
Mattress hose disconnected
Connect hose connectors and lock them
in place
1148139-H
37