Invacare MA95Z Owners Manual - Page 24

Hold Button MA95Z Power Unit, Firm/Soft Buttons, Mode Button MA90Z Power Unit, Low Air Loss Mode,

Page 24 highlights

microAIR® MA90 and MA95 Series 3. Use the Function button to move the pointer to the times or attributes available for the selected function. 4. Use the Firmer or Softer buttons to change the times or attributes. 5.1.7 Hold Button (MA95Z Power Unit) Low Air Loss Mode 1. Press the Hold button G to pause the mattress in the current position. The mattress will remain in this position until the Hold button is pressed again. 5.1.8 Firm/Soft Buttons 1. Select comfort pressure settings by pressing Firm H or Soft I buttons. • Soft Button-Pressing this button reduces the pressure in the mattress. • Firm Button-Pressing this button increases the pressure in the mattress. The patient comfort pressure ranges from Soft (8 ±4 mmHg) to Firm (32 ±4 mmHg). The Comfort Control LED displays the patient comfort pressure levels and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight. If the patient's weight to height ratio is above average, increase the pressure setting by approximately 20%. MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 8 to 32. 5.1.9 Mode Button (MA90Z Power Unit) Press the Mode button J to select the Max Inflate Mode or the Low Air Loss mode. 24 In this mode, the mattress goes into on-demand low air loss relief mode. Max Inflate Mode In this mode, the mattress inflates rapidly to maximum firmness (pressurized to 35 ±5 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting. • It takes 5-10 minutes for the MA90Z mattress to inflate fully (inflation time depends on size of mattress). • Max Inflate mode can be manually disengaged by pressing the Max Inflate button. This will deactivate the Max Inflate LED. • It is recommended that Max Inflate setting be used during patient ingress/egress, patient wound care, patient turning or patient cleaning. 5.1.10 Max Inflate Button (MA95Z Power Unit) Press the Max Inflate button K to select the max inflate mode and inflate the mattress rapidly to maximum firmness (pressurized to 35 ± 5 mmHg). A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active. After 30 minutes, the Max Inflate mode deactivates and the control unit defaults to the previous setting. • It takes 25-60 seconds for the MA95Z mattress to inflate fully (inflation time depends on size of mattress). 1148139-H

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40

microAIR® MA90 and MA95 Series
3.
Use the Function button to move the pointer to the times or attributes available for the selected function.
4.
Use the Firmer or Softer buttons to change the times or attributes.
5.1.7
Hold Button (MA95Z Power Unit)
1.
Press the Hold button
G
to pause the mattress in
the current position. The mattress will remain in this
position until the Hold button is pressed again.
5.1.8
Firm/Soft Buttons
1.
Select comfort pressure settings by pressing Firm
H
or
Soft
I
buttons.
Soft Button—Pressing this button reduces the
pressure in the mattress.
Firm Button—Pressing this button increases the
pressure in the mattress.
The patient comfort pressure ranges from Soft
(8 ±4 mmHg) to Firm (32 ±4 mmHg).
The
Comfort Control LED displays the patient comfort
pressure levels and provides a guide to the
caregiver to set approximate comfort pressure
level depending on the patient weight. If the
patient’s weight to height ratio is above average,
increase the pressure setting by approximately
20%. MA90Z models display levels from 0 to
9 and MA95Z models display the mmHg from
8 to 32.
5.1.9
Mode Button (MA90Z Power Unit)
Press the Mode button
J
to select the Max Inflate Mode
or the Low Air Loss mode.
Low Air Loss Mode
In this mode, the mattress goes into on-demand low air
loss relief mode.
Max Inflate Mode
In this mode, the mattress inflates rapidly to maximum
firmness (pressurized to 35 ±5 mmHg). A series of beeps
sound every three minutes as a reminder that the Max
Inflate mode is active. After 30 minutes, the Max Inflate
mode deactivates and the control unit defaults to the
previous setting.
It takes 5–10 minutes for the MA90Z mattress to inflate
fully (inflation time depends on size of mattress).
Max Inflate mode can be manually disengaged by
pressing the Max Inflate button. This will deactivate
the Max Inflate LED.
It is recommended that Max Inflate setting be used
during patient ingress/egress, patient wound care,
patient turning or patient cleaning.
5.1.10
Max Inflate Button (MA95Z Power Unit)
Press the Max Inflate button
K
to select the max inflate
mode and inflate the mattress rapidly to maximum firmness
(pressurized to 35 ± 5 mmHg). A series of beeps sound every
three minutes as a reminder that the Max Inflate mode is
active. After 30 minutes, the Max Inflate mode deactivates
and the control unit defaults to the previous setting.
It takes 25–60 seconds for the MA95Z mattress to
inflate fully (inflation time depends on size of mattress).
24