7 SETUP/MAINTENANCE 7.6 Removing/Installing the Batteries From/Into the Wheelchair Removing the Front Shroud/Battery Retention Bracket and Rear Shroud ƽ CAUTION Place the wheelchair in a well ventilated area where work can be performed without risking damage to carpeting or floor covering. For this procedure, refer to FIGURE 1 on page 83. 1. Verify the joystick On/Off switch is in the Off position. 2. Remove the three mounting screws that secure the rear shroud to the wheelchair. 3. Remove the rear shroud from the wheelchair. The front shroud and battery retention bracket are removed from and installed onto the wheelchair frame as a single unit. 4. Remove the two mounting screws securing the front shroud/battery retention bracket to the wheelchair. 5. Wheelchairs with Footboards Only: Remove the mounting screw and washer securing the footboard support to the wheelchair frame. 6. Lift up to unhook the battery retention bracket from the wheelchair frame and remove the front shroud/battery retention bracket from the wheelchair. 7. Disconnect the controller from the batteries at the rear of the wheelchair. 8. Remove the batteries. Removing the Batteries from Wheelchair on page 84. Invacare® TDX®SI Series 82 Part No 1154294