Campbell Scientific 253 253-L and 257-L (Watermark 200) Soil Matric Potential - Page 27
Calculate 0.018 * dT
View all Campbell Scientific 253 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 27 highlights
253-L and 257-L Soil Matric Potential Sensors ;{CR10X} *Table 1 Program 01: 1.0000 Execution Interval (seconds) ;Measure soil temperature with 107 sensor 1: Temp (107) (P11) 1: 1 Reps 2: 1 SE Channel 3: 1 Excite all reps w/E1 4: 1 Loc [ Tsoil_C ] 5: 1.0 Multiplier 6: 0.0 Offset ;Measure 257 block resistance 2: AC Half Bridge (P5) 1: 1 Reps 2: 14 250 mV Fast Range 3: 2 SE Channel 4: 2 Excite all reps w/Exchan 2 5: 250 mV Excitation 6: 2 Loc [ kOhms ] 7: 1 Multiplier 8: 0 Offset ;Convert Half Bridge reading to kOhms 3: BR Transform Rf[X/(1-X)] (P59) 1: 1 Reps 2: 2 Loc [ kOhms ] 3: 1 Multiplier (Rf) ;Calculate dT = T -21 4: Z=X+F (P34) 1: 1 X Loc [ Tsoil_C ] 2: -21 F 3: 4 Z Loc [ CorFactr ] ;Calculate (0.018 * dT) 5: Z=X*F (P37) 1: 4 X Loc [ CorFactr ] 2: 0.018 F 3: 4 Z Loc [ CorFactr ] ;Calculate (1 - (0.018 * dT)) 6: Z=X+F (P34) 1: 4 X Loc [ CorFactr ] 2: -1 F 3: 4 Z Loc [ CorFactr ] 7: Z=X*F (P37) 1: 4 2: -1 3: 4 X Loc [ CorFactr ] F Z Loc [ CorFactr ] 21