Campbell Scientific 253 253-L and 257-L (Watermark 200) Soil Matric Potential - Page 32

End of measurement and processing loop, Turn off multiplexer

Page 32 highlights

253-L and 257-L Soil Matric Potential Sensors 9: Z=X*F (P37) 1: 6 -- 2: 0.0106 3: 6 -- X Loc [ WP_kPa_1 ] F Z Loc [ WP_kPa_1 ] 10: Z=F x 10^n (P30) 1: 34.21 F 2: 0 n, Exponent of 10 3: 16 Z Loc [ Const_1 ] 11: Z=X-Y (P35) 1: 16 2: 1 -- 3: 16 X Loc [ Const_1 ] Y Loc [ T107_C_1 ] Z Loc [ Const_1 ] 12: Z=X+Y (P33) 1: 6 -- X Loc [ WP_kPa_1 ] 2: 16 Y Loc [ Const_1 ] 3: 6 -- Z Loc [ WP_kPa_1 ] 13: Z=X*F (P37) 1: 6 -- 2: 1.062 3: 6 -- X Loc [ WP_kPa_1 ] F Z Loc [ WP_kPa_1 ] 14: Z=X-Y (P35) 1: 6 -- 2: 11 -- 3: 6 -- X Loc [ WP_kPa_1 ] Y Loc [ kOhms_1 ] Z Loc [ WP_kPa_1 ] 15: Z=F x 10^n (P30) 1: 1.306 F 2: -2 n, Exponent of 10 3: 17 Z Loc [ Const_2 ] 16: Z=X*Y (P36) 1: 6 -- X Loc [ WP_kPa_1 ] 2: 17 Y Loc [ Const_2 ] 3: 6 -- Z Loc [ WP_kPa_1 ] 17: Z=X/Y (P38) 1: 11 -- 2: 6 -- 3: 6 -- X Loc [ kOhms_1 ] Y Loc [ WP_kPa_1 ] Z Loc [ WP_kPa_1 ] ;End of measurement and processing loop 18: End (P95) ;Turn off multiplexer 19: Do (P86) 1: 51 Set Port 1 Low 26

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36

253-L and 257-L Soil Matric Potential Sensors
9:
Z=X*F (P37)
1:
6
--
X Loc [ WP_kPa_1
]
2:
0.0106
F
3:
6
--
Z Loc [ WP_kPa_1
]
10:
Z=F x 10^n (P30)
1:
34.21
F
2:
0
n, Exponent of 10
3:
16
Z Loc [ Const_1
]
11:
Z=X-Y (P35)
1:
16
X Loc [ Const_1
]
2:
1
--
Y Loc [ T107_C_1
]
3:
16
Z Loc [ Const_1
]
12:
Z=X+Y (P33)
1:
6
--
X Loc [ WP_kPa_1
]
2:
16
Y Loc [ Const_1
]
3:
6
--
Z Loc [ WP_kPa_1
]
13:
Z=X*F (P37)
1:
6
--
X Loc [ WP_kPa_1
]
2:
1.062
F
3:
6
--
Z Loc [ WP_kPa_1
]
14:
Z=X-Y (P35)
1:
6
--
X Loc [ WP_kPa_1
]
2:
11
--
Y Loc [ kOhms_1
]
3:
6
--
Z Loc [ WP_kPa_1
]
15:
Z=F x 10^n (P30)
1:
1.306
F
2:
-2
n, Exponent of 10
3:
17
Z Loc [ Const_2
]
16:
Z=X*Y (P36)
1:
6
--
X Loc [ WP_kPa_1
]
2:
17
Y Loc [ Const_2
]
3:
6
--
Z Loc [ WP_kPa_1
]
17:
Z=X/Y (P38)
1:
11
--
X Loc [ kOhms_1
]
2:
6
--
Y Loc [ WP_kPa_1
]
3:
6
--
Z Loc [ WP_kPa_1
]
;End of measurement and processing loop
18:
End (P95)
;Turn off multiplexer
19:
Do (P86)
1:
51
Set Port 1 Low
26