Campbell Scientific 4WFBS1K 4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Ter - Page 33

Lead Length Resistance per 100 feet.

Page 33 highlights

4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM) 11: Z=X-Y (P35) ; 1: 1 -- X Loc [ mVPerVG01 ] 2: 17 -- Y Loc [ mVPerVZ01 ] 3: 33 Z Loc [ Vr_1 ] 12: Z=X*F (P37) 1: 33 X Loc [ Vr_1 ] 2: -2 F 3: 34 Z Loc [ One_2Vr ] 13: Z=X+F (P34) 1: 34 X Loc [ One_2Vr ] 2: 1000 F 3: 34 Z Loc [ One_2Vr ] 14: Z=X/Y (P38) 1: 33 X Loc [ Vr_1 ] 2: 34 Y Loc [ One_2Vr ] 3: 35 Z Loc [ Vr_1_2Vr ] 15: Z=X/Y (P38) 1: 35 X Loc [ Vr_1_2Vr ] 2: 36 -- Y Loc [ AdjGF01 ] 3: 52 -- Z Loc [ uStrain01 ] 16: Z=X*Y (P36) 1: 52 -- X Loc [ uStrain01 ] 2: 68 Y Loc [ Number4e3 ] 3: 52 -- Z Loc [ uStrain01 ] 17: End (P95) 18: If Flag/Port (P91) 1: 12 Do if Flag 2 is High 2: 30 Then Do 19: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 1 Interval (same units as above) 3: 2 Call Subroutine 2 ; 20: End (P95) *Table 2 Program 02: 0 *Table 3 Subroutines 1: Beginning of Subroutine (P85) ; 1: 1 Subroutine 1 2: Do (P86) ; 1: 11 Set Flag 1 High 3: Z=F (P30) ; 1: 2.5 F ; 2: 0 Exponent of 10 3: 69 Z Loc [ LeadOhms ] 4: Bulk Load (P65) ; 1: 3.0 F ; 2: 4.4 F ; 3: 8 F ; 4: 12 F ; 5: 14 F ; 6: 19 F ; 7: 15 F ; 8: 13 F ; 9: 70 Loc [ LeadFt01 ] Subtract zeroed value from measurement. Outputs data to FinalStorage. Measures ZeroOffset strain reading and sets GF Setup so Subroutine does not get called again. Lead Length Resistance per 100 feet. 0.025 Ohms/Foot for 24 gauge copper stranded wire. Load lead length of the gages in 100s of feet: Gage01 Gage02 Gage03 Gage04 Gage05 Gage06 Gage07 Gage08 27

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46

4WFBS120, 4WFBS350, 4WFBS1K
4 Wire Full Bridge Terminal Input Modules (TIM)
11:
Z=X-Y (P35) ;
Subtract zeroed value from measurement.
1: 1
-- X Loc [ mVPerVG01 ]
2: 17
-- Y Loc [ mVPerVZ01 ]
3: 33
Z Loc [ Vr_1
]
12:
Z=X*F (P37)
1: 33
X Loc [ Vr_1
]
2: -2
F
3: 34
Z Loc [ One_2Vr
]
13:
Z=X+F (P34)
1: 34
X Loc [ One_2Vr
]
2: 1000
F
3: 34
Z Loc [ One_2Vr
]
14:
Z=X/Y (P38)
1: 33
X Loc [ Vr_1
]
2: 34
Y Loc [ One_2Vr
]
3: 35
Z Loc [ Vr_1_2Vr
]
15:
Z=X/Y (P38)
1: 35
X Loc [ Vr_1_2Vr
]
2: 36
-- Y Loc [ AdjGF01
]
3: 52
-- Z Loc [ uStrain01 ]
16:
Z=X*Y (P36)
1: 52
-- X Loc [ uStrain01 ]
2: 68
Y Loc [ Number4e3 ]
3: 52
-- Z Loc [ uStrain01 ]
17:
End (P95)
18:
If Flag/Port (P91)
1: 12
Do if Flag 2 is High
2: 30
Then Do
19:
If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 1
Interval (same units as above)
3: 2
Call Subroutine 2 ;
Outputs data to FinalStorage.
20:
End (P95)
*Table 2
Program
02:
0
*Table 3 Subroutines
1:
Beginning of Subroutine (P85) ;
Measures ZeroOffset strain reading and sets GF
1: 1
Subroutine 1
2:
Do (P86) ;
Setup so Subroutine does not get called again.
1: 11
Set Flag 1 High
3:
Z=F (P30) ;
Lead Length Resistance per 100 feet.
1: 2.5
F ;
0.025 Ohms/Foot for 24 gauge copper stranded wire.
2: 0
Exponent of 10
3: 69
Z Loc [ LeadOhms
]
4:
Bulk Load (P65) ;
Load lead length of the gages in 100s of feet:
1: 3.0
F
;
Gage01
2: 4.4
F ;
Gage02
3: 8
F ;
Gage03
4: 12
F ;
Gage04
5: 14
F ;
Gage05
6: 19
F ;
Gage06
7: 15
F ;
Gage07
8: 13
F ;
Gage08
9: 70
Loc [ LeadFt01
]
27