Campbell Scientific 4WFBS1K 4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Ter - Page 7
WFBS120, 4WFBS350, 4WFBS1K, Full Bridge Terminal Input Modules TIM
![]() |
View all Campbell Scientific 4WFBS1K manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 7 highlights
4WFBS120, 4WFBS350, 4WFBS1K 4 Wire Full Bridge Terminal Input Modules (TIM) 1. Function The 4WFBS120, 4WFBS350, and 4WFBS1K Terminal Input Modules (TIM) complete a full Wheatstone bridge for a single strain gage or other sensor that acts as a single variable resistor. The difference between the three models is in the resistor that matches the nominal resistance of a 120 ohm, 350 ohm, or 1000 ohm quarter bridge strain gage. It can also be used to complete the back half of a Wheatstone bridge for use in a ¼ bridge strain circuit (1 active element) using a dummy gage, or in a ½ bridge strain circuit (2 active elements). FIGURE 1-1. Terminal Input Module with CR1000 2. Specifications 2:1 Resistive Divider Resistors: Ratio tolerance @ 25 °C: Ratio temperature coefficient: Power rating per element: 1 kΩ/1 kΩ ±0.01% 0.5 ppm/°C (-55°C to 85°C) 0.1 W @ 70°C Completion Resistor: 120, 350, or 1000 Ω Tolerance @ 25 °C: ±0.01% Temperature coefficient: ±0.8 ppm °C-1 (-55°C to 85°C) Power rating: 0.25 W @ 70°C 1
![](/manual_guide/products/campbell-scientific-4wfbs120-4wfbs120-4wfbs350-4wfbs1k-4-wire-bridge-terminal-input-modules-cfa4f38/7.png)