Campbell Scientific WINDSONIC1 WINDSONIC 2-D Sonic Wind Sensors - Page 31
SDI-12 Transparent Mode on an Edlog array-based
View all Campbell Scientific WINDSONIC1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 31 highlights
WindSonic Two-Dimensional Sonic Anemometer • To activate the SDI-12 Transparent Mode, on Control Port p, enter pX and press the key. The datalogger will respond with "entering SDI12". If any invalid SDI-12 command is issued, the datalogger will exit the SDI-12 Transparent Mode. • To query the WindSonic4 for its current SDI-12 address, enter the command ?!. The WindSonic4 will respond with the current SDI-12 address. • To change the SDI-12 address, enter the command aAb!; where a is the current address from the above step and b is the new address. The WindSonic4 will change its address and the datalogger will exit the SDI12 Transparent Mode. • To activate the SDI-12 Transparent Mode on Control Port p, enter pX and press the key. Verify the new SDI-12 address by entering the "?!" command. The WindSonic4 will respond with the new address. • To exit the SDI-12 Transparent Mode, enter ∗. FIGURE 7-1. SDI-12 Transparent Mode on an Edlog array-based datalogger (CR10X) using control port 8 and changing the SDI-12 address from 0 to 1 23