Campbell Scientific WINDSONIC1 WINDSONIC 2-D Sonic Wind Sensors - Page 33
CR200(X)-series Datalogger, SDI12Recorder
View all Campbell Scientific WINDSONIC1 manuals
Add to My Manuals
Save this manual to your list of manuals |
Page 33 highlights
WindSonic Two-Dimensional Sonic Anemometer FIGURE 7-2. SDI-12 Transparent Mode on an Edlog table-based datalogger using control port 8 and changing the SDI-12 address from 0 to 1 7.5.3 CR200(X)-series Datalogger • Connect a single WindSonic4 to the datalogger using Control Port C1/SDI12 as described in Section 7.2.3, SDI-12 Wiring, and download a datalogger program that does not contain the SDI12Recorder() instruction. • In the LoggerNet Toolbar, navigate to and activate the Test|Terminal Emulator ... menu. The "Terminal Emulator" window will open. In the Select Device menu, located in the lower left hand side of the window, select the station. • Click on the Open Terminal button. If communications between the datalogger and PC are successful, the red bar located in the upper left hand side of the window will turn green. • Press the key until the datalogger responds with the "CR200(X)>" prompt (FIGURE 7-3). • To query the WindSonic4 for its current SDI-12 address, press the key, at the "CR200(X)>" prompt enter the command "SDI12>?!", and 25