Fluke 123 FE 123 & 124 Users Manual - Page 46

Using the 10:1 Probe for High Frequency, Measurements., Probe Attenuation.

Page 46 highlights

Fluke 123/124 Users Manual The reading now shows the risetime from 10%-90% of the trace amplitude and the voltage at the cursors in relation to the zero icon (-). See Figure 1-21. ~ Turn off the cursors. 38 Using the 10:1 Probe for High Frequency Measurements. Fluke 124 is supplied with a model VP40 10:1 Probe. Use of this Probe is recommended when you measure high frequency signals in circuits with a high impedance. The loading of the circuit by a 10:1 Probe is much lower than that of a 1:1 Shielded Test Lead. The following aspects must be observed when using a 10:1 Probe: Probe Attenuation. The Probe attenuates the signal 10 times. Proceed as follows to adapt the Test Tool's voltage readout to this attenuation. The example below is for a Probe connected to input B: Open the Scope menu. Open the Probes menu. Select PROBE on B ... Press ENTER Select 10:1 V

  • 1
  • 2
  • 3
  • 4
  • 5
  • 6
  • 7
  • 8
  • 9
  • 10
  • 11
  • 12
  • 13
  • 14
  • 15
  • 16
  • 17
  • 18
  • 19
  • 20
  • 21
  • 22
  • 23
  • 24
  • 25
  • 26
  • 27
  • 28
  • 29
  • 30
  • 31
  • 32
  • 33
  • 34
  • 35
  • 36
  • 37
  • 38
  • 39
  • 40
  • 41
  • 42
  • 43
  • 44
  • 45
  • 46
  • 47
  • 48
  • 49
  • 50
  • 51
  • 52
  • 53
  • 54
  • 55
  • 56
  • 57
  • 58
  • 59
  • 60
  • 61
  • 62
  • 63
  • 64
  • 65
  • 66
  • 67
  • 68
  • 69
  • 70
  • 71
  • 72
  • 73
  • 74
  • 75
  • 76
  • 77
  • 78
  • 79
  • 80
  • 81
  • 82
  • 83
  • 84
  • 85
  • 86

Fluke 123/124
Users Manual
38
The reading now shows the risetime from 10%-90% of the
trace amplitude and the voltage at the cursors in relation
to the zero icon (-). See Figure 1-21.
Turn off the cursors.
Using the 10:1 Probe for High Frequency
Measurements.
Fluke 124 is supplied with a model VP40 10:1 Probe. Use
of this Probe is recommended when you measure high
frequency signals in circuits with a high impedance. The
loading of the circuit by a 10:1 Probe is much lower than
that of a 1:1 Shielded Test Lead.
The following aspects must be observed when using a
10:1 Probe:
Probe Attenuation.
The Probe attenuates the signal 10 times. Proceed as
follows to adapt the Test Tool’s voltage readout to this
attenuation. The example below is for a Probe connected
to input B:
±
Open the Scope menu.
²
Open the Probes menu.
³
Select PROBE on B ...
´
Press ENTER
µ
Select 10:1 V