Fluke 123/124 Users Manual Open the SCOPE INPUTS menu. Open the PROBES submenu. Highlight PROBE AC ADJUST. Open the PROBE AC ADJUST submenu. Highlight ADJUST 10:1 PROBE. A square wave appears on the screen. 48 Adjust the trimmer screw in the probe housing to give an optimum square wave. Return to normal mode.